• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE JETSTREAM® XC ATHERECTOMY CATHETER; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE JETSTREAM® XC ATHERECTOMY CATHETER; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Model Number 112266-001
Device Problems Complete Blockage (1094); Aspiration Issue (2883); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/15/2016
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.(b)(4).Device evaluated by mfr: it is indicated that the device will not be returned for evaluation.A review of the batch history, historical trending, and similar complaint trending review for the product family will be conducted.If there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
 
Event Description
It was reported that the device stopped spinning and became clogged.A 2.4mm jetstream® xc atherectomy catheter was selected over a non-bsc guidewire to treat a calcified lesion in the lower leg.The physician actively stopped aspiration to take a picture to evaluate the progress.Upon repositioning the catheter in the vessel, it was alleged that it didn¿t sound right upon starting the device again so he stopped using it and removed it from the body.Outside of the body he tried the catheter and noticed it wasn¿t spinning.The device was used for a total of 6 minutes before it stopped spinning.Additionally, it was alleged that the catheter could have become clogged.The procedure was completed with a different device.No patient complications were reported and the patient's condition was reported as fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JETSTREAM® XC ATHERECTOMY CATHETER
Type of Device
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5573092
MDR Text Key42506589
Report Number2134265-2016-02678
Device Sequence Number1
Product Code MCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130637
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 03/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2017
Device Model Number112266-001
Device Catalogue NumberPV41340
Device Lot Number18763203
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/15/2016
Initial Date FDA Received04/13/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/06/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
COOK GUIDEWIRE
-
-