• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PROGRAMMING SOFTWARE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PROGRAMMING SOFTWARE Back to Search Results
Model Number MODEL 250
Device Problems Intermittent Continuity (1121); Device Displays Incorrect Message (2591); Communication or Transmission Problem (2896)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/21/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported by a company representative that a physician's programmer was having intermittent communication issues.The company representative did not recall the exact error message observed, but the issues would occasionally resolve by manipulating the usb serial data cable.On the date of the report the company representative was unable to resolve the issue.Follow-up to the company representative revealed that a new usb serial data cable fixed the issue.The suspect usb serial data cable has not been received to-date.Additional relevant information has not been received to-date.
 
Event Description
The usb serial data cable was received by the manufacturer (b)(6) 2016.Analysis was completed on the returned usb serial data cable (b)(6) 2016 and the report was verified.The cause of the issue was associated with a disconnected wire connection.Once the wire was soldered onto the printed circuit board, no further anomalies were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING SOFTWARE
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5573297
MDR Text Key43172987
Report Number1644487-2016-00788
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 03/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Other Device ID NumberVERSION 11.0.5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2016
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 03/21/2016
Initial Date FDA Received04/13/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/03/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/27/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-