During the patient's full vns revision, a new 3mm lead was implanted.After the lead was implanted, high impedance was observed but the surgeon only attempted to reposition the electrodes on the patient nerve and would not test to see about relieving the back-pressure in the vns generator header or to see if the lead pin was fully inserted into the generator header.A new 2mm lead was implanted which also resulted in high impedance.It was noted the company representative again suggested checking the lead pin insertion, but the surgeon initially dismissed the suggestion.Another surgeon then noted that the lead pin didn't appear to be fully inserted.At that time the pin was re-inserted and the diagnostic results were within normal limits.It was reported the surgeon later noted after the case that he may have prematurely replaced that lead and apologized for the haste decision.Attempts for additional relevant information have been unsuccessful to date.
|