Information was received from a company representative regarding a patient who was implanted with a neurostimulator for urinary dysfunction/sacral nerve stim and gastrointestinal/pelvic floor.It was reported that they had no issue inserting the lead into first or second implantable neurostimulator (ins).They performed electrode impedance test at 1.5v and 360pw.The impedance read high 4000 ohms.They reran impedance test several times on the call at difference settings.They initially saw c-3 with question marks and all bipolar pairs, except 2-3 greater than 4000 ohms.They ran impedance again and question marks cleared and only the following pairs were showing greater than 4,000 ohms: 0-1, 0-2, 0-3, 1-2.It was indicated the ins was being used for testing motor response.They programmed and used previous effective pairs or pairs that were currently high at rate 14, pw 210us, cycling set at 3s on and 3 s off, no soft start.They increased amplitude with physician programer(pp).They saw motor response with the following pairs at the noted amplitudes 0-1 - 0.7v 0-2 - 2.1v 0-3 - 3.2v 1-2 - 1.6v.It was indicated that since high impedances pairs were providingmotor response and high impedances might resolve post-op, it would be appropriate to close patient.
|