The device was in use for treatment.A review of the device history record could not be performed as the lot number of the cable is unknown.The extension cable was not returned to perform an analysis.As a result, the allegation against the cable cannot be confirmed.This cable is 100% final inspected for correct colored part, measured and verified for overall length, checked for damage and insulation voids, verified that the silicone strain relief extends at least for a length of 2 cm outside the female connector, the silicone strain relief is checked so that it fits the wire snugly, strain relieves are checked for damage and gaps, and a pull test is done to verify strength of the soldered joint.The cable is also 100 percent inspected for continuity and for proper connector function.The instructions for use (ifu) informs the user that the cable can be re-sterilized by oscor eto gas sterilization a maximum of two times.Precautions are provided to the user: do not connect any cable model to a/c power source; connection of exposed pins to a/c power source may pose a risk of serious injury or death.Corrective and preventive action has been initiated to address this issue.
|