• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OSCOR INC. MEDTRONIC REUSABLE EXTENSION CABLE; ELECTRODE CABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OSCOR INC. MEDTRONIC REUSABLE EXTENSION CABLE; ELECTRODE CABLE Back to Search Results
Model Number 53912
Device Problem Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/13/2015
Event Type  malfunction  
Manufacturer Narrative
The device was in use for treatment.A review of the device history record could not be performed as the lot number of the cable is unknown.The extension cable was not returned to perform an analysis.As a result, the allegation against the cable cannot be confirmed.This cable is 100% final inspected for correct colored part, measured and verified for overall length, checked for damage and insulation voids, verified that the silicone strain relief extends at least for a length of 2 cm outside the female connector, the silicone strain relief is checked so that it fits the wire snugly, strain relieves are checked for damage and gaps, and a pull test is done to verify strength of the soldered joint.The cable is also 100 percent inspected for continuity and for proper connector function.The instructions for use (ifu) informs the user that the cable can be re-sterilized by oscor eto gas sterilization a maximum of two times.Precautions are provided to the user: do not connect any cable model to a/c power source; connection of exposed pins to a/c power source may pose a risk of serious injury or death.Corrective and preventive action has been initiated to address this issue.
 
Event Description
The customer reported to their distributor that the extension cable broke near the red/black connectors that are plugged into the external pulse generator.There was no report of any patient adverse effects from this event.This customer perceives the cable as too thin and that it breaks with normal use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDTRONIC REUSABLE EXTENSION CABLE
Type of Device
ELECTRODE CABLE
Manufacturer (Section D)
OSCOR INC.
3816 desoto blvd.
palm harbor FL 34683 1618
Manufacturer (Section G)
OSCOR INC.
3816 desoto blvd.
palm harbor FL 34683 1618
Manufacturer Contact
jan flegeau
3816 desoto blvd.
palm harbor, FL 34683-1618
7279372511
MDR Report Key5579310
MDR Text Key42750164
Report Number1035166-2016-00071
Device Sequence Number1
Product Code IKD
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K070926
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,user faci
Reporter Occupation Other
Type of Report Initial
Report Date 03/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Health Care Professional
Device Model Number53912
Device Catalogue Number53912
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/21/2016
Initial Date FDA Received04/14/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-