BAXTER HEALTHCARE - IRVINE INFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number J2C1702
Device Problem Hole In Material (1293)
Patient Problem No Patient Involvement (2645)
Event Date 03/24/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was received for evaluation with approximately 10 ml of fluid still in the bladder.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Microscopic inspection did not reveal any holes or defects on the tubing.A functional leak test was also performed and the device operated within specification.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that there was a hole in the tubing of a small volume infusor.This was observed before use of the device.There was no patient involvement.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
bldg 3
irvine CA 92614
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
bldg 3
irvine CA 92614
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5579317
MDR Text Key42753120
Report Number1416980-2016-07278
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 04/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date05/01/2018
Device Catalogue NumberJ2C1702
Device Lot Number15E039
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/01/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/24/2016
Initial Date FDA Received04/14/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/20/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-