• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION PRECISION®; SPINAL CORD STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION PRECISION®; SPINAL CORD STIMULATOR Back to Search Results
Model Number SC-2218-50
Device Problems Bent (1059); Fracture (1260); High impedance (1291); Kinked (1339)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/30/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Additional suspect medical device component involved in the event: model #: sc-2218-50, serial #: (b)(4), description: linear st lead, 50cm, sc-2218-50 (sn (b)(4)).The complaint has been confirmed.Visual (microscope) and x-ray inspection of the lead revealed that all of the cables were completely broken at the bent/kinked location of the lead.The bent/kinked locations are 1 cm from both sides of the set screw mark of the clik anchor.There are no exposed cables at the fracture locations.Sc-2218-50 (sn (b)(4)) the complaint has been confirmed.Visual (microscope) and x-ray inspection of the lead revealed that 2 cables were fractured at the bent/kinked location of the lead.The bent/kinked location is 1 cm from the set screw mark of the clik anchor.No cables are exposed at the fracture site.
 
Event Description
A report was received that the patient's leads had high impedances.The patient underwent a revision procedure wherein the leads were replaced.No device malfunction was suspected.The patient was doing well postoperatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION®
Type of Device
SPINAL CORD STIMULATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key5582948
MDR Text Key42851095
Report Number3006630150-2016-00912
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
030017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/30/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/01/2015
Device Model NumberSC-2218-50
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/30/2015
Initial Date FDA Received04/15/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/30/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age46 YR
-
-