(b)(4).Additional suspect medical device component involved in the event: model #: sc-2218-50, serial #: (b)(4), description: linear st lead, 50cm, sc-2218-50 (sn (b)(4)).The complaint has been confirmed.Visual (microscope) and x-ray inspection of the lead revealed that all of the cables were completely broken at the bent/kinked location of the lead.The bent/kinked locations are 1 cm from both sides of the set screw mark of the clik anchor.There are no exposed cables at the fracture locations.Sc-2218-50 (sn (b)(4)) the complaint has been confirmed.Visual (microscope) and x-ray inspection of the lead revealed that 2 cables were fractured at the bent/kinked location of the lead.The bent/kinked location is 1 cm from the set screw mark of the clik anchor.No cables are exposed at the fracture site.
|