• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. CDI BLOOD PARAMETER MONITORING SYSTEM 500; MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. CDI BLOOD PARAMETER MONITORING SYSTEM 500; MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500 Back to Search Results
Model Number 500AV
Device Problems Fluid/Blood Leak (1250); Failure to Run on Battery (1466)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/05/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The reported issue was confirmed.During the laboratory evaluation, the bpm unit did not last the required 25 minutes of run-time on standby battery.The product surveillance technician (pst) charged the bpm unit's 12 volt (v) direct current (d/c) standby battery overnight to a minimum of eight plus hours.The bpm unit was placed in operate mode the following day, alternating current (a/c) power removed, the battery lasted less than 30 seconds and turned off immediately.The pst's inspection of the bpm unit's 12v battery revealed that it has leaked and fluid has been released onto rear chassis.The bpm unit's 12v d/c battery was defective; leaking and doesn't hold a charge to operate monitor.The product will be sent to service to be brought to manufacturers specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the battery back-up was not working on the blood parameter monitor (bpm).As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CDI BLOOD PARAMETER MONITORING SYSTEM 500
Type of Device
MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key5584737
MDR Text Key43585052
Report Number1828100-2016-00259
Device Sequence Number1
Product Code DRY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial
Report Date 04/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number500AV
Device Catalogue Number500AV
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/07/2016
Initial Date FDA Received04/18/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-