Device Problems Leak/Splash (1354); Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 03/21/2016
Event Type  malfunction  
Manufacturer Narrative
A 3rd party service representative performed a checkout of the equipment and confirmed the reported complaint.The bag/vent micro-switch was replaced, and the unit was returned to service.
 
Event Description
The hospital reported a failure where the bag/vent switch leaked and failed to switch to mechanical ventilation during pre-use checkout.There is no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AESPIRE 7100
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi 21402 8
CH  214028
Manufacturer (Section G)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi 21402 8
CH   214028
Manufacturer Contact
stephanie cass
3000 n. grandview boulevard
waukesha, WI 53188-1696
MDR Report Key5590578
MDR Text Key43123913
Report Number9710602-2016-00014
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K000706
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 04/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/21/2016
Initial Date FDA Received04/20/2016
Date Device Manufactured07/20/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-