• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THERAKOS, INC THERAKOS CELLEX PHOTOPHERESIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THERAKOS, INC THERAKOS CELLEX PHOTOPHERESIS SYSTEM Back to Search Results
Lot Number D366-KIT
Device Problem Obstruction of Flow (2423)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/17/2016
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.A batch record review of kit lot d366 was conducted.There were no nonconformances associated with this lot.The lot met release requirements.The uvadex lot number was not provided as it was not administered.However, a review of all uvadex lots manufactured since january 2013 was performed.No trends or nonconformances related to the complaint were noted.Trends were reviewed for complaint categories, clot observed and access issues.No trends were detected for these complaint categories.The assessment is based on information available at the time of the investigation.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Complaints are monitored through tracking and trending.If a trend is detected, further investigation will be conducted through the capa/continuous improvement process.(b)(4).
 
Event Description
The customer called to report that both the patient's access and the procedural kit "clotted off" during the procedure.The customer stated that this occurred just before buffy coat collection.The customer reported that she could not remember which alarms occurred, but that the centrifuge bowl had stopped.The customer stated that when she looked in the bowl, the volume in the bowl appeared to be clotted.The customer reported that they aborted the procedure with no blood returned to the patient.The customer stated that the patient received a blood transfusion in order to account for the blood loss.The customer reported that both she and the clinicians believed that the clotting was related to a patient issue, and was not an instrument or kit issue.The customer confirmed that an anticoagulant was used and that it was spiked correctly into the anticoagulant spike line.The customer stated that the patient was discharged from the clinic in stable condition and returned for another photopheresis procedure on (b)(6) 2016.The kit was not returned for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERAKOS CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX
Manufacturer (Section D)
THERAKOS, INC
hampton NJ
Manufacturer (Section G)
HARMAC MEDICAL PRODUCTS
2201 bailey ave
buffalo NY 14211
Manufacturer Contact
megan vernak
53 frontage road
hampton, NJ 08827
MDR Report Key5592131
MDR Text Key44188777
Report Number2523595-2016-00109
Device Sequence Number1
Product Code LNR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Nurse
Type of Report Initial
Report Date 03/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Expiration Date11/01/2017
Device Lot NumberD366-KIT
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/21/2016
Initial Date FDA Received04/20/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age23 YR
Patient Weight50
-
-