• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES DR DISPOSABLE PRESSURE TRANSDUCER; TRANSDUCER, PRESSURE, CATHETER TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES DR DISPOSABLE PRESSURE TRANSDUCER; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number T001691M
Device Problems Fluid/Blood Leak (1250); Output above Specifications (1432)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/30/2016
Event Type  malfunction  
Manufacturer Narrative
The lot number was not provided; therefore, a review of the manufacturing records could not be completed.Unfortunately, we did not receive the sample for examination in our product evaluation laboratory.Without return of the product, we are unable to perform a complete investigation into the root cause of the reported event.At this time, we cannot confirm the reported issue.
 
Event Description
As reported, this dpt (disposable pressure transducer) provided systolic pressures values 30 mm/hg higher than the values provided by a non-edwards, non-invasive pressure cuff (arm band).The usual process was followed by checking the line, pulse flushing using the reservoir with saline solution to remove potential occlusions at the cannula.However, the saline solution squirted out of the port.Finally, the patient was treated using a flotrac sensor.There was no allegation of patient injury.Unfortunately, the sample is not available for evaluation as it was discarded by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE PRESSURE TRANSDUCER
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
carretera sanchez, km 18 1/2
parque ind. de itabo
haina. san cristobal
DR 
Manufacturer (Section G)
EDWARDS LIFESCIENCES DR
carretera sanchez, km 18 1/2
parque ind. de itabo
haina. san cristobal
DR  
Manufacturer Contact
brian hurd-servin
1 edwards way
irvine, CA 92614
9492506423
MDR Report Key5592453
MDR Text Key44023822
Report Number2015691-2016-01304
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 03/31/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberT001691M
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/31/2016
Initial Date FDA Received04/20/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-