It was reported that diagnostic tests were taken on the generator prior to implanting of device outside of the pocket and after device was in the pocket.The test before implant showed the battery was not at near end of service condition, but the test after the generator was in the pocket showed that the device was at end of service condition.It was reported that the generator was not cold prior to implant.The physician decided to continue implanting the generator.The manufacturer's or specialist that attended the surgery did not see cautery come in contact with the generator, but the cautery device was lying on the patient when the generator was introduced.The surgeon also submerged the generator in a tray with antibiotic solution prior to implant.The manufacturer investigated the programming data available from surgery.The voltage dropped from 3.287v to 1.884v during the surgery.The initial interrogation (before implant) and system diagnostics were at ifi = no, and the subsequent diagnostics (after implant) measured between 1.884v and 1.988v.The drop in voltage is an indicator that the device likely came in contact with an electro-cautery device or an electrostatic discharge, causing the asic latch-up condition.Electro-cautery is known to potentially cause battery depletion and its usage is warned against in labeling.No known surgical intervention has occurred to date.
|