STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON ASYM PATELLA A35X10; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5551L350
Device Problems Device Operates Differently Than Expected (2913); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Fall (1848); Injury (2348)
Event Date 03/28/2016
Event Type  Injury  
Manufacturer Narrative
Additional information has been requested.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event Description
Revision knee for pain suffered after a fall for a 3 year old knee implant.Doctor removed scar tissue and replaced well fixed patella.Femur and tibia were well fixed and left in place.
 
Manufacturer Narrative
An event regarding revision due to arthrofibrosis involving a triathlon patellar component was reported.The event was not confirmed.Method & results: device evaluation and results: the returned patellar component was visually inspected.One of the pegs has been cut from the patella, with marks evident on the remaining section which is consistent with damaged caused during explantation as pegs are often cut during revision to remove the component from the patella bone.The patellar component otherwise appears unremarkable for a poly component that was implanted and subsequently explanted.Medical records received and evaluation: not performed as no medical information was provided.Device history review: dhr review for the reported lot determined that the device was manufactured and packed to specification.Complaint history review: complaint history review confirmed that there have been no other similar events for the reported lot.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as pre- and post-operative x-rays, operative reports as well as patient history and follow-up notes are needed to complete the investigation for determining a root cause.A capa trend analysis was conducted for the reported failure mode and concluded that arthrofibrosis may result from other factors not necessarily related to the device.No further investigation is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
Revision knee for pain suffered after a fall for a (b)(4) year old knee implant.Doctor removed scar tissue and replaced well fixed patella.Femur and tibia were well fixed and left in place.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON ASYM PATELLA A35X10
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5597761
MDR Text Key43319550
Report Number0002249697-2016-01288
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/28/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/31/2018
Device Catalogue Number5551L350
Device Lot NumberLDJ831
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/28/2016
Initial Date FDA Received04/22/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/30/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/30/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age58 YR
-
-