It was reported that the patient's device heartbeat verification was unable to be performed by the neurologist during a follow up appointment.Programming history is available from date of implant, (b)(6) 2015 to (b)(6) 2016.One system diagnostic test was performed on (b)(6) 2015 and shows normal impedance of 2310 ohms.The battery indicator is ifi - no.On the date of implant, normal mode output current remained programmed off during the heartbeat testing but the magnet and auto-stimulation output current were turned on.Tachycardia detection was programmed on, and heartbeat detection at sensitivity level 1 was successful (60 beats per minute).On (b)(6) 2016, the heartbeat verification was attempted at sensitivity settings 5 and 4 and resulted in heart rate of 77.6 beats per minute.At the end of the visit, the device was programmed to heartbeat sensitivity setting 5, with td on and all output currents on with resulting heart rate as 67.8 beats per minute.A review of device history records for the generator shows that no unresolved non-conformances were found.
|