• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COBAS E-ELECSYS ANTI-CCP; ANTIBODIES, ANTI-CYCLIC CITRULLINATED PEPTIDE (CCP)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COBAS E-ELECSYS ANTI-CCP; ANTIBODIES, ANTI-CYCLIC CITRULLINATED PEPTIDE (CCP) Back to Search Results
Catalog Number 05031656190
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/01/2016
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The customer complained of erroneous results for 1 patient tested for anti-ccp.The erroneous results were reported outside of the laboratory.The initial anti-ccp result from the e411 analyzer was 30.79 u/ml.This result was reported outside of the laboratory.A new sample was obtained 3 days later and run on the e411 analyzer again and the result was "the same." the initial sample was repeated on a different e411 analyzer at a different customer site and the result was "the same." the initial sample was repeated on an abbott architect instrument and the result was 1.5 u/ml.A new sample was obtained and the result from a siemens system was <1.2 u/ml.No adverse event occurred.The e411 analyzer serial number was not provided.
 
Manufacturer Narrative
A specific root cause could not be identified.Additional information for further investigation was requested but was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS E-ELECSYS ANTI-CCP
Type of Device
ANTIBODIES, ANTI-CYCLIC CITRULLINATED PEPTIDE (CCP)
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key5612405
MDR Text Key43854729
Report Number1823260-2016-00523
Device Sequence Number1
Product Code NHX
Combination Product (y/n)N
Reporter Country CodePK
PMA/PMN Number
K081338
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 05/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2016
Device Catalogue Number05031656190
Device Lot Number188366
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/07/2016
Initial Date FDA Received04/27/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/11/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-