A hospital in (b)(6), reported that a set up which involved a fisher & paykel healthcare (fph) mr850 respiratory humidifier, ltv1150 ventilator and pulmonetics/airlife breathing circuit was allegedly over heating and more than normal condensate was found in the breathing circuit.They further reported that the ventilation was disrupted and that the patient's heart rate and respiratory rate increased and a mild burn was noted on the circumference of the tracheostomy stoma.The patient's condition was reported to be stable and that nystatin lotion was used on the reported burn.
|
(b)(4).The complaint mr850jhu respiratory humidifier was not returned to fisher & paykel healthcare for investigation.Our analysis is accordingly based on the additional information provided by the homecare provider (i.E.Absolute respiratory care), and our knowledge of the product.The homecare provider reported that the subject mr850 humidifier, which was initially used with the ltv1150 ventilator and pulmonetics/airlife breathing circuit, was tested at their office and no fault was found during inspection.It was also reported that the condensation observed in the subject breathing circuit was due to the electric fans inside the room that were constantly turned on while the therapy was being administered.The homecare provider also commented that the environmental condition at the patient's house was unstable.When reporting the complaint, the homecare provider confirmed that the patient was in stable condition and that the subject mr850 humidifier was replaced with the hc550 respiratory humidifier.The fph hc500 series respiratory humidifiers are intended to deliver optimal humidity to pediatric and adult patients in their homes.The homecare provider confirmed that the subject mr850 humidifier was functioning correctly during inspection and that the reported condensation was influenced by environmental conditions.Condensate in the humidification system, although not preferred, is an expected side effect of heated pass-over humidification systems in many conditions, and may vary between light misting to water droplets that form on the wall of cool breathing circuit tubing.The amount of condensate in the ventilation system is influenced by multiple set-up and environmental factors.The product technical manual (ptm) that accompany the mr850 respiratory humidifier state that the device is "designed for use in hospital intensive care units".Further, it also states "the use of breathing circuits, chambers or other accessories which are not approved by fisher & paykel healthcare may impair performance or compromise safety.".
|
A hospital in (b)(6) reported that a set up, which involved a fisher & paykel healthcare (fph) mr850 respiratory humidifier, ltv1150 ventilator and pulmonetics/airlife breathing circuit, was allegedly over heating and more than normal condensate was found in the breathing circuit.They further reported that the ventilation was disrupted and that the patient's heart rate and respiratory rate increased and a mild burn was noted on the circumference of the tracheostomy stoma.The patient's condition was reported to be stable and that nystatin lotion was used on the reported burn.
|