Model Number XPO100
Device Problem No Display/Image (1183)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Should additional information become available a supplemental record will be filed.
 
Event Description
Dealer states the unit has a bad display.
 
Manufacturer Narrative
Product was returned for evaluation.The return fields in oracle state: portable concentrators.Sieve bed, saturated.Pcb, other/defective.Controls, other/defective.Complaint was confirmed.The underlying cause could not be determined after reviewing the documentation in this investigation.
 
Event Description
Dealer states the unit has a bad display.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPO2 PORTABLE CONCENTRATOR 9153648166
Type of Device
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
INVACARE FLORIDA OPERATIONS
2101 east lake mary blvd
sanford FL 32773
Manufacturer (Section G)
INVACARE FLORIDA OPERATIONS
2101 east lake mary blvd
sanford FL 32773
Manufacturer Contact
kevin guyton
one invacare way
elyria, OH 44035
8003336900
MDR Report Key5619389
MDR Text Key44612393
Report Number1031452-2016-02048
Device Sequence Number1
Product Code CAW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 05/04/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberXPO100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/18/2016
Initial Date FDA Received04/29/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/17/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-