• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) NAV THORACIC PROBE TIP; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) NAV THORACIC PROBE TIP; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9734680
Device Problems Bent (1059); Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/04/2016
Event Type  malfunction  
Manufacturer Narrative
Patient information was not made available from the site.Device lot number not available.Device manufacturing date is dependent on lot number, therefore, unavailable.Return requested.Replacement thoracic probe shipped to site 04/05/2016.No parts have been received by manufacturer for analysis.No further issues have been reported.
 
Event Description
A medtronic representative reported that, while in a spine procedure, the site's straight thoracic probe tip was bent.No further details regarding the damage, or how it occurred, were provided.The procedure was continued using a second probe.The surgeon completed the procedure with the use of the navigation system.Delay in therapy was 30 minutes.There was no impact on patient outcome.
 
Manufacturer Narrative
(b)(4).Medtronic investigation of returned suspect thoracic probe finds that the tip of the probe was slightly bent, and a small nick at the tip.Testing also found impact marks at the back end of the probe.The fit with the navlock tracker was difficult as well.The hardware investigation found that the reported event was related to a hardware issue.The reported event was confirmed to be caused by physical damage.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAV THORACIC PROBE TIP
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
kristianna bilan
826 coal creek circle
louisville, CO 80027-9710
7208902338
MDR Report Key5624797
MDR Text Key44367470
Report Number1723170-2016-00668
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K124004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 06/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9734680
Device Lot Number111021
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/04/2016
Initial Date FDA Received05/03/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/01/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/21/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age50 YR
-
-