A customer in great britain reported two incidences where their thinprep 5000 processor with autoloader etched a wrong vial id onto a slide.The customer put the vial through again and could not reproduce the error.The instrument did not present any error to the operator.The vials were put through the machine again and the slide number was etched correctly.Therefore, there was no delay in patient diagnosis.Hologic technical service (ts) requested for a picture of the vials, but the customer wouldn't send any.Instrument is operational.Fse needs to be dispatched.(b)(4) field service engineer (fse) confirmed but unable to reproduce error.Performed all required setups per technical documentation.Performed (b)(4).Instrument operational.Although no patients needed to be recalled as a result of this incident this is a reportable event since the instrument did not perform as intended.
|