Catalog Number 254500060
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/27/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The attune pin jack broke.
 
Manufacturer Narrative
This complaint remains under investigation.Depuy will notify the fda of the results of the investigation once it has been completed.
 
Manufacturer Narrative
Examination of the returned device confirms the reported event of breakage.The received 254500060 attune pin puller was forwarded to depuy material science for review.On the basis of these observations, the fracture of the pin grasper of the pin jack was caused by ductile overload, indicative of a bending force applied with the handle with respect to an engaged pin exceeding the ultimate strength of the material.Based on the root cause of overload corrective action is not indicated.Continue to monitor via (b)(4).Depuy considers the investigation closed at this time.Should the additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PIN PULLER
Type of Device
KNEE INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key5633581
MDR Text Key44465897
Report Number1818910-2016-18652
Device Sequence Number1
Product Code HWB
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 05/04/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254500060
Device Lot NumberPG231419
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/04/2016
Initial Date FDA Received05/05/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received05/05/2016
06/28/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/15/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Weight108
-
-