• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES DR PRESSURE MONITORING SET; TRANSDUCER, PRESSURE, CATHETER TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES DR PRESSURE MONITORING SET; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number T001650A
Device Problems Break (1069); Detachment of Device or Device Component (2907)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 04/08/2016
Event Type  malfunction  
Manufacturer Narrative
The device evaluation is anticipated.However the complaint could not be confirmed without the completion of the product evaluation.A supplemental report will be forthcoming with the evaluation results.
 
Event Description
It was reported that the dpt (disposable pressure transducer) tubing detached from the reservoir.There was no allegation of patient injury.Device was available for evaluation.
 
Manufacturer Narrative
We received one bifurcated iv set for examination.As received the iv tubing was detached from the bond joint with the drip chamber.No apparent indication of bonding material was observed on either the iv tubing or drip chamber bond areas.The instructions indicate to bond the iv tubing and drip chamber with bonding solvent (mixture of cyclohexanone and mek).No other damage or defect was observed from the returned iv set.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint.A review of the manufacturing records indicated that the product met specifications upon release.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSURE MONITORING SET
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
DR 
Manufacturer (Section G)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
DR  
Manufacturer Contact
brian hurd-servin
1 edwards way
irvine, CA 92614
9492506423
MDR Report Key5634685
MDR Text Key45309853
Report Number2015691-2016-01491
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K142749
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 04/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/02/2017
Device Model NumberT001650A
Device Lot Number60174243
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/14/2016
Initial Date FDA Received05/05/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/05/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/03/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-