(b)(4).Date of initial report: (b)(6) 2016.One used ls1037 device was received, and evaluation of the returned sample could not confirm the reported issue with activation.Testing of the device found that it functions within all related specifications, and would seal on simulated tissue without issues.However, a separate non-contributing issue of damaged wire insulation was found.The manufacturing process has been reviewed and nothing found that would cause this type of damage.The device history records for this lot were also reviewed and no potentially contributing entries found.Damage of this type is consistent with scraping the jaws against the sharp edge of a trocar during insertion or removal of the device.The ifu included with this product cautions users to carefully insert and withdraw instruments from trocars to avoid possible damage to the devices and/or injury to the patient.
|
The customer reported that during a procedure, the device would not activate.Upon receipt of the device, covidien identified that a piece of the wiring insulation next to the jaws was missing and not returned.The customer has confirmed that the missing piece did not fall within the patient cavity.No patient injury occurred.
|