Catalog Number 425-00 |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/02/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Qn#(b)(4).A visual, functional and dimensional inspection of the product involved in the complaint could not be conducted since the device was not returned at the time of this report.The device history record (dhr) review did not show issues related to the complaint.The device was manufactured on 11/19/2008.A document assessment (fmea-08-037) was conducted and no changes were required.The reported complaint cannot be confirmed since the device sample is not available to perform a proper investigation, determine the root cause and assign corrective actions.If the device sample becomes available at a later date, this complaint will be updated accordingly.
|
|
Event Description
|
The event is reported as: the unit is overheating.Unknown when alleged defect was detected.
|
|
Manufacturer Narrative
|
(b)(4).The sample was returned for evaluation.A visual exam was performed and there were no defects observed.Functional testing was performed and the unit was connected to 110vac.The unit passed the initial power connect test and navigated through the power-on self-test (p.O.S.T.) with no issues.During the temperature display accuracy test the unit displayed the correct temperatures and properly alarmed in the high temperature scenario.The unit was put on functional bench testing for one hour.The unit ran uninterrupted for that time at set temperature of 37 degrees c and heavy rainout.Based on the investigation performed, the reported complaint could not be confirmed.Functional testing did not reveal any operational anomalies.
|
|
Event Description
|
The event is reported as: the unit is overheating.Unknown when alleged defect was detected.
|
|
Search Alerts/Recalls
|