• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. (CRM-KISTA) MERLIN PROGRAMMER; PACEMAKER PROGRAMMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. (CRM-KISTA) MERLIN PROGRAMMER; PACEMAKER PROGRAMMER Back to Search Results
Model Number 3650
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Patient Involvement (2645)
Event Date 02/12/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Analysis description: final analysis found excess dust in the power supply, which led to a rise in temperature within the programmer.The damage found likely contributed to the reported burn of the device.
 
Event Description
It was reported that prior to an implant procedure, the programmer exhibited a burning smell and no longer turned on.The unit was no longer used and was replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN PROGRAMMER
Type of Device
PACEMAKER PROGRAMMER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC. (CRM-KISTA)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC. (CRM-KISTA)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
ulla strindlund
box 7051
isafjordsgatan 15
kista SE-16-407
SW   SE-16407
4684744043
MDR Report Key5640573
MDR Text Key44721105
Report Number3010215456-2016-03051
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 02/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3650
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/29/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/12/2016
Initial Date FDA Received05/09/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-