• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION PYXIS ES; MEDICATION DISPENSING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION PYXIS ES; MEDICATION DISPENSING Back to Search Results
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Information (3190)
Event Date 04/20/2016
Event Type  malfunction  
Event Description
Issue with pyxis es machine failing drawers leaving medications and emergency supplies unavailable.A scheduled monthly generator load test took place on 04/20/2016 at 0500.Backup batteries were previously installed (04/19/2016) onto each pyxis es machine in the hospital in anticipation of the load test.Upon utility interruption, two pyxis es machines located in labor and delivery (pyxis es machine wmc ld2) and nicu (pyxis es machine (b)(6) nicu2) experienced multiple drawer failure.Additional individual pocket failures were experienced in critical care 2, critical care 3, observation unit, emergency room east, and emergency room north.A total of 390 drawers failed seconds after the utility interruption occurred.The majority of the failures were experienced in labor and delivery and nicu.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PYXIS ES
Type of Device
MEDICATION DISPENSING
Manufacturer (Section D)
CAREFUSION
MDR Report Key5640725
MDR Text Key44791697
Report NumberMW5062141
Device Sequence Number1
Product Code JQP
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 05/05/2016
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received05/05/2016
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other;
-
-