BOSTON SCIENTIFIC - MARLBOROUGH ADVANTAGE FIT SYSTEM; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR STRESS URINARY INCONTINENCE, RETR
|
Back to Search Results |
|
Model Number M0068502110 |
Device Problem
Difficult to Remove (1528)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/07/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).The device has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that an advantage fit was used during a sling case performed on (b)(4) 2016.According to the complainant, during the procedure and inside the patient, the blue sleeve scrunched up and was retained inside the patient.The incision was enlarged and the physician dissected down to retrieve the plastic sheath and pulled it up with a kelly clamp.The procedure was completed with another advantage fit.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
|
|
Manufacturer Narrative
|
Visual analysis of the returned advantage fit device confirmed that the dilator was bunched and with two holes.The delivery device was not returned.The device most likely encountered anatomical or procedural factors which limited its performance, therefore, the most probable root cause classification for the reported failure is operational context.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution. a search of the complaint database confirmed that no similar complaints exist for the specified batch. the investigation concluded that this complaint is associated with a product that meets design and manufacture specifications but due to anatomical/procedural factors encountered during the procedure, performance of the device was limited.
|
|
Event Description
|
It was reported to boston scientific corporation that an advantage fit was used during a sling case performed on (b)(6) 2016.According to the complainant, during the procedure and inside the patient, the blue sleeve scrunched up and was retained inside the patient.The incision was enlarged and the physician dissected down to retrieve the plastic sheath and pulled it up with a kelly clamp.The procedure was completed with another advantage fit.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
|
|