Model Number M0068502110
Device Problem Difficult to Remove (1528)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/07/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).The device has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that an advantage fit was used during a sling case performed on (b)(4) 2016.According to the complainant, during the procedure and inside the patient, the blue sleeve scrunched up and was retained inside the patient.The incision was enlarged and the physician dissected down to retrieve the plastic sheath and pulled it up with a kelly clamp.The procedure was completed with another advantage fit.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Manufacturer Narrative
Visual analysis of the returned advantage fit device confirmed that the dilator was bunched and with two holes.The delivery device was not returned.The device most likely encountered anatomical or procedural factors which limited its performance, therefore, the most probable root cause classification for the reported failure is operational context.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution. a search of the complaint database confirmed that no similar complaints exist for the specified batch. the investigation concluded that this complaint is associated with a product that meets design and manufacture specifications but due to anatomical/procedural factors encountered during the procedure, performance of the device was limited.
 
Event Description
It was reported to boston scientific corporation that an advantage fit was used during a sling case performed on (b)(6) 2016.According to the complainant, during the procedure and inside the patient, the blue sleeve scrunched up and was retained inside the patient.The incision was enlarged and the physician dissected down to retrieve the plastic sheath and pulled it up with a kelly clamp.The procedure was completed with another advantage fit.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANTAGE FIT SYSTEM
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR STRESS URINARY INCONTINENCE, RETR
Manufacturer (Section D)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key5641119
MDR Text Key44736008
Report Number3005099803-2016-01321
Device Sequence Number1
Product Code OTN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020110
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date05/20/2018
Device Model NumberM0068502110
Device Catalogue Number850-211
Device Lot NumberML00003003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/11/2016
Initial Date FDA Received05/09/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/02/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/09/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age42 YR
-
-