• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER, INC. (JUAREZ) LASSO® 2515 NAV ECO VARIABLE CATHETER; CATHETER, ELECTRODE RECORDING OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER, INC. (JUAREZ) LASSO® 2515 NAV ECO VARIABLE CATHETER; CATHETER, ELECTRODE RECORDING OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number D-1343-02-S
Device Problems Bent (1059); Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/19/2016
Event Type  malfunction  
Manufacturer Narrative
Since the product was not returned for analysis, no product failure analysis can be conducted and no determination of possible contributing factors could be made.As such, the investigation will be closed.If the complaint device is received in the future we will reopen the complaint and perform the investigation as appropriate.The device history record (dhr) was reviewed and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.(b)(4).
 
Event Description
It was reported that a patient underwent a paroxysmal atrial fibrillation procedure with a lasso 2515 nav eco variable catheter and the deflection became stuck as the shaft of the catheter was bent and stuck.There were no wires exposed on the shaft of the catheter and there was no difficulty removing the catheter from the patient through the sheath.The catheter was replaced and the issue resolved.The procedure was completed with no patient consequence.This event is mdr reportable because the lasso was stuck in a deflected position which could potentially cause patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LASSO® 2515 NAV ECO VARIABLE CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
BIOSENSE WEBSTER, INC. (JUAREZ)
circuito interior norte #1820
parque industrial salvacar
juarez, chihuahua 32599
MX  32599
Manufacturer (Section G)
BIOSENSE WEBSTER, INC. (JUAREZ)
circuito interior norte #1820
parque industrial salvacar
juarez, chihuahua 32599
MX   32599
Manufacturer Contact
joaquin kurz
33 technology drive
irvine, CA 92618
9497893837
MDR Report Key5651263
MDR Text Key46261178
Report Number9673241-2016-00327
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
PMA/PMN Number
K113213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2018
Device Model NumberD-1343-02-S
Device Catalogue NumberD134302
Device Lot Number17371096L
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/19/2016
Initial Date FDA Received05/12/2016
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/03/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-