Reportable based on device analysis completed on 19apr2016.It was reported that catheter balloon leak occurred.The target lesion was located in the coronary artery.A 15/3.00 flextome cutting balloon catheter was selected for use.During procedure, the physician inserted the catheter and inflate it up in the vessel; however, it was noted that there was a drip on the balloon that was going up at nominal.The physician then deflated the system and brought the balloon out.The balloon was again inflated and the physician noted that there was a slight leaking on the device.The device was completely removed from the patient.The procedure was completed with an unspecified balloon catheter.No patient complications were reported and the patient's status is fine.However, device analysis revealed a balloon pinhole 0.9mm proximal from the proximal end of the distal markerband.
|