• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY ICL (IMPLANTABLE COLLAMER LENS); INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY ICL (IMPLANTABLE COLLAMER LENS); INTRAOCULAR LENS Back to Search Results
Model Number VICMO13.2
Device Problems Break (1069); Improper or Incorrect Procedure or Method (2017); Torn Material (3024)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/05/2016
Event Type  malfunction  
Manufacturer Narrative
This product is manufactured in the u.S.But not marketed in the u.S.(b)(4).
 
Event Description
The reporter indicated the surgeon inserted a 13.2mm vicmo13.2 implantable collamer lens, -13.50 diopter, in the patient's left eye (os) and the lens tore/broke during injection/delivery into the eye.The lens was removed with no apparent patient injury.The lens was exchanged for another same model lens and the problem was resolved.
 
Manufacturer Narrative
Additional information: device evaluation: product evaluation found that the lens was returned dry in the lens case/vial.Visual inspection found that the haptic was broken and bent.Corrected data: (b)(4).The reporter states that an error in lens loading caused the tear in the lens.Conclusion code: the reporter states that a use error was made during lens loading.Claim # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICL (IMPLANTABLE COLLAMER LENS)
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
michelle andres
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key5655185
MDR Text Key45192149
Report Number2023826-2016-00702
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date07/31/2018
Device Model NumberVICMO13.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/26/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/14/2016
Initial Date FDA Received05/13/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/10/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/26/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age29 YR
-
-