Model Number MMT-523NAL
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Hyperglycemia (1905)
Event Date 04/25/2016
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
Customer called to report damage to the insulin pump.Customer reported that the back of the pump including the drive support cap is coming off.Customer's blood glucose reading was 170 mg/dl.Advised discontinuation of the insulin pump and revert to back up plan per health care professional.Product is being returned and replaced.
 
Manufacturer Narrative
The insulin pump was received with detached end cap, missing end cap sticker and cracked reservoir tube lip.The drive support disk was inspected and no anomaly was noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key5660184
MDR Text Key45409840
Report Number3004209178-2016-48801
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
PMA/PMN Number
P980022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-523NAL
Device Catalogue NumberMMT-523NAL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/11/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/25/2016
Initial Date FDA Received05/17/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/20/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age31 YR
Patient Weight73
-
-