Model Number MMT-523NAL |
Device Problem
Device Dislodged or Dislocated (2923)
|
Patient Problem
Hyperglycemia (1905)
|
Event Date 04/25/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Event Description
|
Customer called to report damage to the insulin pump.Customer reported that the back of the pump including the drive support cap is coming off.Customer's blood glucose reading was 170 mg/dl.Advised discontinuation of the insulin pump and revert to back up plan per health care professional.Product is being returned and replaced.
|
|
Manufacturer Narrative
|
The insulin pump was received with detached end cap, missing end cap sticker and cracked reservoir tube lip.The drive support disk was inspected and no anomaly was noted.
|
|