Catalog Number AR-13991N
Device Problems Break (1069); Difficult to Remove (1528)
Patient Problem Foreign Body In Patient (2687)
Event Date 04/19/2016
Event Type  Injury  
Manufacturer Narrative
Patient demographics (age at time of event, date of birth, gender, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.The device was requested for evaluation but was not returned therefore the complainant's event could not be verified.The cause of the event could not be determined from the information available and without device evaluation.Device history record review revealed nothing relevant to this event.The typical cause for this type of event would be the use of excessive force to pass the needle through thick or hard tissue or hitting bone with the needle.There is a label on the device warning the user against re-sterilizing, reusing, hitting bone or use of excessive force as these may result in needle breakage or patient injury.This is the first complaint of this type for this part/lot combination.The potential cause(s) of this event will be communicated to the event reporter.If additional relevant information is received, a follow-up report will be submitted.Device requested but not yet received.
 
Event Description
It was reported that the tip of device broke inside of patient's shoulder joint while undergoing arthroscopic shoulder surgery.X-ray was taken that identified tip of device as being inside patient but unable to be removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUREFIRE SCORPION NEEDLE
Type of Device
INSTRUMENT, MANUAL, SURGICAL, GENERAL USE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath, sr. mdr analyst.
1370 creekside boulevard
naples, FL 34108-1945
8009337017
MDR Report Key5660759
MDR Text Key45335689
Report Number1220246-2016-00222
Device Sequence Number1
Product Code MDM
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 04/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/01/2020
Device Catalogue NumberAR-13991N
Device Lot Number1200053Z
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/25/2016
Initial Date FDA Received05/17/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-