• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH UNIVERSAL DRIVER SHAFT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH UNIVERSAL DRIVER SHAFT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 2107-1015
Device Problems Break (1069); Fracture (1260); Material Fragmentation (1261); Material Integrity Problem (2978)
Patient Problems No Consequences Or Impact To Patient (2199); Device Embedded In Tissue or Plaque (3165)
Event Date 04/20/2016
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event Description
During a right total hip primary, the tip of the screwdriver broke off and was unable to be extracted.No adverse consequence to the patient.
 
Manufacturer Narrative
An event regarding crack/fracture involving a trident driver was reported.The event was confirmed.Method & results: device evaluation and results: the hexalobular tip of was broken off.The broken piece was not included with the returned driver shaft.Medical records received and evaluation: not performed as there were no medical records provided.Device history review: review of the device history records indicates devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been 7 other events for the lot referenced.The three events had the screw tip broken off during usage.Conclusions: during visual inspection it was noted that the hexalobular driver tip fractured.The tip threads are also deformed which indicates excessive force by the user.
 
Event Description
During a right total hip primary, the tip of the screwdriver broke off and was unable to be extracted.No adverse consequence to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL DRIVER SHAFT
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5661382
MDR Text Key45428928
Report Number0002249697-2016-01626
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2107-1015
Device Lot NumberF7E15591
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/20/2016
Initial Date FDA Received05/17/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/26/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/12/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age61 YR
Patient Weight104
-
-