• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK STATUS+; CT STATUS+

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK STATUS+; CT STATUS+ Back to Search Results
Catalog Number 10379675
Device Problems Device Emits Odor (1425); Smoking (1585)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/18/2016
Event Type  malfunction  
Manufacturer Narrative
Customer has been requested to return instrument for further evaluation.The root cause for the event is unknown.
 
Event Description
Customer stated that she plugged unit in and it had a strange smell to it so she unplugged unit and let it sit for approximately an hour then plugged unit back in and unit had strange smell again and it started smoking.Customer confirmed that there was no report of injury due to this event.
 
Manufacturer Narrative
Instrument was returned for investigation.Investigation of the instrument determined that the failure was cause due to a failure of d16 on the main pcb.D16 is the overvoltage protector diode, which blows if the voltage applied to instrument is too high.This complaint is not confirmed.Customer is operational with replacement instrument.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITEK STATUS+
Type of Device
CT STATUS+
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
steven andberg
2 edgewater drive
norwood, MA 02062
7812693655
MDR Report Key5663261
MDR Text Key46662534
Report Number1217157-2016-00059
Device Sequence Number1
Product Code JIL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/26/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number10379675
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/26/2016
Initial Date FDA Received05/18/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/26/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage N
Patient Sequence Number1
-
-