Catalog Number 10379675 |
Device Problems
Device Emits Odor (1425); Smoking (1585)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/18/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Customer has been requested to return instrument for further evaluation.The root cause for the event is unknown.
|
|
Event Description
|
Customer stated that she plugged unit in and it had a strange smell to it so she unplugged unit and let it sit for approximately an hour then plugged unit back in and unit had strange smell again and it started smoking.Customer confirmed that there was no report of injury due to this event.
|
|
Manufacturer Narrative
|
Instrument was returned for investigation.Investigation of the instrument determined that the failure was cause due to a failure of d16 on the main pcb.D16 is the overvoltage protector diode, which blows if the voltage applied to instrument is too high.This complaint is not confirmed.Customer is operational with replacement instrument.
|
|