MAUDE Adverse Event Report: HILL-ROM BATESVILLE CLINITRON BED; BED FLOTATION THERAPY POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE CLINITRON BED; BED FLOTATION THERAPY POWERED Back to Search Results
Model Number 0800
Device Problem Loss of Power (1475)
Patient Problem Injury (2348)
Event Date 04/26/2016
Event Type  Injury  
Manufacturer Narrative
The patient developed a new wound on her backside.(b)(6) receives home health care monday-friday and the wound is being treated with dressings.Preexisting wounds.The patient is in the bed full time.The hill-rom technician found the sheet, filter, tray, microspheres, control panel, diffuser board, and blower needed to be replaced.Per the hill-rom user manual, poor fluidization; if fluidization is sluggish or uneven, notify your hill-rom representative.Fluidization is affected by the following: room temperature, humidity, the amount of materials, such as fluid, cells or cellular debris, which has escaped from the blood vessels and has been deposited in tissues or on tissue surfaces, restricted air circulation from blankets on the bed.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.Development of pressure ulcers is multifactorial and cannot be only attributed to performance of the surface.Risk factors include protein-calorie malnutrition, microclimate (skin wetness caused by sweating or incontinence), diseases that reduce blood flow to the skin, such as arteriosclerosis, or diseases that reduce the sensation in the skin, such as paralysis or neuropathy.Position changes are key to pressure sore prevention and treatment.These changes need to be frequent, repositioning needs to avoid stress on the skin, and body positions need to minimize the risk of pressure on vulnerable areas.No further information is available on the repair of the bed at this time.If any additional relevant information is identified following completion of the repair, the additional relevant information will be submitted in a supplemental report.
 
Event Description
Hill-rom received a report from the account stating the bed shut down and the patient developed a new wound.The bed was located at the account.There was a patient injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITRON BED
Type of Device
BED FLOTATION THERAPY POWERED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
john cummings
1069 state route 46 east
batesville, IN 47006
812931-286
MDR Report Key5670299
MDR Text Key45625436
Report Number1824206-2016-00225
Device Sequence Number1
Product Code IOQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK964223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 04/26/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number0800
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/26/2016
Initial Date FDA Received05/20/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age70 YR
-
-