Model Number 840 |
Device Problem
Poor Quality Image (1408)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The service engineer (se) evaluated the ventilator and replaced the graphic user interface (gui) printed circuit board (pcb) and backlight inverter printed circuit boards (pcbs).The software level was updated.The ventilator passed self-testing.
|
|
Event Description
|
It was reported that the ventilator's top display was hard to read.The ventilator was not on a patient at the time of the event.
|
|
Manufacturer Narrative
|
The device was repaired and the reported issue was isolated to interface between the device and the failed component.If information is provided in the future, a supplemental report will be issued.
|
|