• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON VENTILATOR TUBING SET,LONG LENGTH; CIRCUIT, BREATHING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON VENTILATOR TUBING SET,LONG LENGTH; CIRCUIT, BREATHING Back to Search Results
Catalog Number 1613
Device Problem Detachment Of Device Component (1104)
Patient Problem No Patient Involvement (2645)
Event Date 05/11/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A visual, functional and dimensional inspection of the product involved in the complaint could not be conducted since the product was not provided.The device history record of batch number (b)(4) has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint.No non conformance reports were originated for the lot in question that can be associated to the complaint reported.Dhr shows that the product was assembled & inspected according to our specifications.No corrective action can be established at this moment since the device sample is not available for evaluation.Customer complaint cannot be confirmed based only on the information provided, to perform an investigation and determine the source of defect reported it is necessary to evaluate the sample involved in this complaint.If the device sample becomes available at a later date this complaint will be updated.
 
Event Description
The customer alleges that the expiratory side of the circuit keeps coming off the expiratory filter.The alleged issue was detected during set-up.
 
Manufacturer Narrative
(b)(4).The sample was returned for evaluation.A visual exam was performed and no defects were observed.The ventilator filter connectors were checked for size compliance and were found to be within specification.The tubing was physically attached to the ventilator filter that was returned along with the sample.It should be noted that as the tubing connector is being pushed into a secure position on the male end connector of the filter, resistance can be detected the further the female end of the tubing connector is pushed.Forceful pushing onto the male end of the ventilator filter will accomplish a tight seal that is not easily detached.Based on the investigation performed, the reported complaint could not be confirmed.There were no issues found with the returned device.
 
Event Description
The customer alleges that the expiratory side of the circuit keeps coming off the expiratory filter.The alleged issue was detected during set-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON VENTILATOR TUBING SET,LONG LENGTH
Type of Device
CIRCUIT, BREATHING
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key5676354
MDR Text Key45809624
Report Number3004365956-2016-00220
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number1613
Device Lot Number74C1601455
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/20/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/12/2016
Initial Date FDA Received05/24/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/16/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/09/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-