It was reported that a vns patient underwent full replacement surgery on (b)(6) 2016.It was reported that the lead was replaced due to lead discontinuity.The patient's vns system was tested upon connection of the new lead to the new generator and system diagnostics returned impedance results within normal limits with 2878 ohms.Further information was received indicating that the generator was prophylactically replaced.It was reported that the explanted devices will not be returned to the manufacturer; therefore, no analysis could be performed.Review of manufacturing records confirmed that both, the generator and the lead passed all functional tests prior to distribution.No additional relevant information was provided to date.
|