Model Number 10-60-00 |
Device Problem
Failure to Pump (1502)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 04/14/2016 |
Event Type
No Answer Provided
|
Manufacturer Narrative
|
There was no patient involvement.Sorin group (b)(4) manufactures the s3 roller pump.The incident occurred in bad (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the rotor of the s3 roller pump did not turn during priming.There was no patient involvement.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
|
|
Event Description
|
Sorin group (b)(4) received a report that the rotor of the s3 roller pump did not turn during priming.There was no patient involvement.
|
|
Manufacturer Narrative
|
Livanova (b)(4) manufactures the s3 roller pump.The incident occurred in bad (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).The livanova field service representative was dispatched to the customers¿ facility to investigate.The service representative identified that the pump did not turn without lid and variolock.Sap service quotation was not accepted because the customer wanted to purchase a new machine.No additional information.If additional information is received it will be evaluated for reportability as required.A review of the dhr did not identify any manufacturing deviations or non-conformities relevant to the reported issue.The root cause was determined to be related to defective components.Livanova has determined that a capa is not required, as there was no patient harm reported.Livanova will continue to monitor this issue for trends.
|
|