MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S3 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S3 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-60-00
Device Problem Failure to Pump (1502)
Patient Problem No Patient Involvement (2645)
Event Date 04/14/2016
Event Type  No Answer Provided  
Manufacturer Narrative
There was no patient involvement.Sorin group (b)(4) manufactures the s3 roller pump.The incident occurred in bad (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the rotor of the s3 roller pump did not turn during priming.There was no patient involvement.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Event Description
Sorin group (b)(4) received a report that the rotor of the s3 roller pump did not turn during priming.There was no patient involvement.
 
Manufacturer Narrative
Livanova (b)(4) manufactures the s3 roller pump.The incident occurred in bad (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).The livanova field service representative was dispatched to the customers¿ facility to investigate.The service representative identified that the pump did not turn without lid and variolock.Sap service quotation was not accepted because the customer wanted to purchase a new machine.No additional information.If additional information is received it will be evaluated for reportability as required.A review of the dhr did not identify any manufacturing deviations or non-conformities relevant to the reported issue.The root cause was determined to be related to defective components.Livanova has determined that a capa is not required, as there was no patient harm reported.Livanova will continue to monitor this issue for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S3 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key5680797
MDR Text Key45962850
Report Number9611109-2016-00303
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K950990
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 12/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-60-00
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/26/2016
Initial Date FDA Received05/25/2016
Supplement Dates Manufacturer Received12/11/2017
Supplement Dates FDA Received12/12/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/18/1998
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-