A dosi-fuser 100ml 2 day infuser was used for a 2 day infusion of fluorouracil.When patient returned to clinic to disconnect dosi-fuser, notified nurse of white powder around restrictor element.Nurse in turn notified pharmacist, who witnessed white powder.Upon inspection, pharmacist and nurse determined the restrictor element may be either cracked or not sealed properly.Nurse wiped the powder from patient's chest and minor skin irritation was noted.During a two week return visit, patient stated the skin became red and irritated, but was at the moment healing and brown in color.Appears to possibly be a chemical burn from the fluorouracil.No further treatment of the area was needed and the patient continued chemotherapy.
|