Model Number BL1UT
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/31/2015
Event Type  Injury  
Manufacturer Narrative
Investigation of this event is in progress.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
It was reported that the lens kept buckling due to unspecified mechanical complications.The lens was explanted and a new lens of a different model was implanted.Additional information has been requested but has not been received.
 
Manufacturer Narrative
The lens was returned to b+l.Visual inspection found a portion of both haptic plates damaged/missing and a tear on the optic.Due to the condition in which the lens was returned, further testing could not be performed.The device history records were reviewed and there were no discrepancies or unusual finding that relate to the reported issue.Based on the information available, the root cause of the reported event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIGN TORIC POSTERIOR CHAMBER IOL
Type of Device
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
BAUSCH + LOMB
rochester NY 14609
Manufacturer (Section G)
BAUSCH + LOMB
10574 acacia street suite d1
rancho cucamonga CA 91730
Manufacturer Contact
sharon spencer
50 technology drive west
irvine, CA 92618
9493985698
MDR Report Key5688671
MDR Text Key46198519
Report Number0001313525-2016-00253
Device Sequence Number1
Product Code MJP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/03/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/28/2019
Device Model NumberBL1UT
Device Lot Number7441409
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/19/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/03/2016
Initial Date FDA Received05/31/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/23/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/12/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age63 YR
-
-