Engineering analysis: the icast was removed from the packaging and inspected to determine the cause of the complaint.The catheter was clean and had no signs of contamination from blood anywhere on the stent or the delivery system.There was a noticeable kink in the catheter shaft just prior to the proximal balloon weld.The stent was properly positioned between the radio opaque marker bands (ro markers) and was not loose on the balloon.The crimped stent diameter was measure and was 2.4mm.This diameter is indicative of a properly crimped stent and matches the quality inspection data attached to this investigation.To determine the functionality of the returned device the delivery system was prepped and the stent deployed to 8atm and then 12atm.Nominal deployment pressure as specified on the product label is 8atm and the rated burst pressure is 12atm.The stent deployed properly and was situated between the two ro marker bands as expected.The device is fully functional.Engineering summary: a full review of the catheter lot history records for the device in question was performed.The records indicate that this lot of catheters passed atrium's final lot qualification testing.This inspection requires that the catheter lot must pass the following: ability of the stent and delivery system to be passed through the labeled introducer sheath; ability to deploy the stent at nominal pressure (8atm); ability to withdraw the deflated balloon catheter back through the labeled introducer sheath; ability of the delivery system to withstand 10 inflate/deflate cycles at the rated burst pressure (12atm) without leaks or failures; balloon burst testing.The balloon must burst over the rated burst pressure specified on the label (12atm); manifold to shaft tensile testing.Samples when tensile tested must not break or separate below 3.37lbs result: all 59 quality inspection samples passed this final inspection without any non-conformances noted during the final lot qualification testing related to the stent.None of the samples tested had any stent movement while passing the device through the introducer sheath.Conclusion: based on the details of the event and the successful lot qualification test data, atrium can find no fault with the device and or lot of stent delivery systems in question.Clinical evaluation: there are several possibilities that can cause stent dislodgement including but not limited to manipulation of the stent prior to use, overuse of the sheath causing failure of the hemostatic valve and a hemostatic valve that is too tight.The instructions for use state that, "special care must be taken not to handle or in any way disrupt the placement of the icast stent on the balloon." insufficient stent securement during access through the sheath may lead to the need for removal of the stent delivery system.This event may be associated with the potential harm of a delay in treatment as well as the need for additional medical or surgical intervention.
|