BOSTON SCIENTIFIC - MARLBOROUGH SYMPHION¿ TISSUE REMOVAL SYSTEM FLUID MANAGEMENT ACCESSORIES; INSUFFLATOR, HYSTEROSCOPIC, FLUID, CLOSED-LOOP RECIRCULATION WITH CUTTER-COAGULA
|
Back to Search Results |
|
Model Number FG-0202 |
Device Problem
Device Inoperable (1663)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/13/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Reported event of pressure sensor failed.Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that a symphion fluid management accessories kit was to be used in the uterus during a hysteroscopic myomectomy procedure performed on (b)(6) 2016.According to the complainant, during preparation, the pressure sensor test failed and the controller displayed an error message instructing to "replace pressure sensor." the procedure was completed using another symphion fluid management accessories kit.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be good.
|
|
Manufacturer Narrative
|
One symphion fluid management accessory device was received for analysis.A functional evaluation of the device noted that it was possible for the pressure sensor connector to be inserter into the receptacle on the controller, and for the pressure sensor to physically connect to the endoscope.However, the controller displayed the following error during setup: "pressure sensor test failed.Replace pressure sensor." a review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.A review of all available information failed to indicate a root cause or probable root cause for this event.
|
|
Event Description
|
It was reported to boston scientific corporation that a symphion fluid management accessories kit was to be used in the uterus during a hysteroscopic myomectomy procedure performed on (b)(6) 2016.According to the complainant, during preparation, the pressure sensor test failed and the controller displayed an error message instructing to "replace pressure sensor." the procedure was completed using another symphion fluid management accessories kit.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be good.
|
|