Model Number N/A |
Device Problem
Bent (1059)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/04/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
Lot numbers tc51452 and tc60020.The device evaluation is anticipated, a follow up report will be sent upon completion of the product evaluation.Report one of two for the same event, see also 3004485144-2016-00096.
|
|
Event Description
|
The sales associate reported screws that were damaged during surgery.The doctor was performing a lateral fusion procedure with telluride system.The set screw was bent as there was difficulty coming down on the rod, and the pedicle screw end tip was bent.
|
|
Manufacturer Narrative
|
The returned device was visually evaluated revealing that the set screw had retained some thread damage.It is likely that the screw was not fully aligned when seated within the tulip head.The complaint is confirmed.The dhr shows that there were no nonconformances or manufacturing issues noted that may have contributed to the event.The reporting rep noted that the patient in the procedure had hard bone meaning this likely contributed to the damages that occurred with the implants.It is also likely that the set screw was misaligned upon insertion, leading to the thread deformation that is seen.
|
|