• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET SPINE - BROOMFIELD SET SCREW 5.5MM ROD TI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET SPINE - BROOMFIELD SET SCREW 5.5MM ROD TI Back to Search Results
Model Number N/A
Device Problem Bent (1059)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/04/2016
Event Type  Injury  
Manufacturer Narrative
Lot numbers tc51452 and tc60020.The device evaluation is anticipated, a follow up report will be sent upon completion of the product evaluation.Report one of two for the same event, see also 3004485144-2016-00096.
 
Event Description
The sales associate reported screws that were damaged during surgery.The doctor was performing a lateral fusion procedure with telluride system.The set screw was bent as there was difficulty coming down on the rod, and the pedicle screw end tip was bent.
 
Manufacturer Narrative
The returned device was visually evaluated revealing that the set screw had retained some thread damage.It is likely that the screw was not fully aligned when seated within the tulip head.The complaint is confirmed.The dhr shows that there were no nonconformances or manufacturing issues noted that may have contributed to the event.The reporting rep noted that the patient in the procedure had hard bone meaning this likely contributed to the damages that occurred with the implants.It is also likely that the set screw was misaligned upon insertion, leading to the thread deformation that is seen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SET SCREW 5.5MM ROD TI
Type of Device
SET SCREW
Manufacturer (Section D)
BIOMET SPINE - BROOMFIELD
310 interlocken parkway
suite 120
broomfield CO 80021
Manufacturer (Section G)
BIOMET SPINE - BROOMFIELD
310 interlocken parkway
suite 120
broomfield CO 80021
Manufacturer Contact
michelle cole
310 interlocken parkway
suite 120
broomfield, CO 80021
3034437500
MDR Report Key5694674
MDR Text Key46397237
Report Number3004485144-2016-00095
Device Sequence Number1
Product Code KWP
Combination Product (y/n)N
PMA/PMN Number
PK100888
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/04/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number7703-1600
Device Lot NumberSEE H10
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/04/2016
Initial Date FDA Received06/02/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/01/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age52 YR
-
-