The explanted devices was received on 06/02/2016.The generator performed according to functional specifications.Analysis of the generator in the pa lab concluded that no abnormal performance or any other type of adverse condition was found.An analysis was performed on the returned lead portions and the reported allegations of; 'fracture of lead(s)' (lead section) was not confirmed.A portion of the lead assembly including the electrodes was not returned for analysis; therefore a complete evaluation could not be performed on the entire lead product.The abraded openings found on the outer silicone tubing and the cut ends that were made during the explanted process, most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids inside the outer silicone tubing.For the observed inner tubing fluid remnants, there was no obvious path for fluid ingress other than the cut ends that were made during the explanted process.With the exception of the observed abraded opening the condition of the returned lead portions is consistent with conditions that typically exist following an explant procedure.No obvious anomalies were noted.The set screw marks found on the lead connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portions were performed, during the visual analysis, and no other discontinuities were identified.Based on the findings in the product analysis lab, there is no evidence to suggest a discontinuity in the returned portions of the device which may have contributed to the stated allegation of 'fracture of lead(s)'.
|