BOSTON SCIENTIFIC - COSTA RICA (COYOL) OPTICROSS¿; CATHETER, ULTRASOUND, INTRAVASCULAR Back to Search Results
Model Number H749518080
Device Problem Kinked (1339)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/24/2016
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr: the device was returned for analysis.The unit returned has the imaging window detached from the lapjoint area.No kinks were found during the testing.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause was unable to be determined.(b)(4).
 
Event Description
Reportable based on device analysis completed on 09may2016.It was reported that shaft kinked occurred.An opticross jp imaging catheter was selected for use.During the using of this opticross, the area from blue changes to transparency shaft got kinked and was unable to be used.The procedure was completed with another of the same device.No patient complications reported and the patient's condition is good.However, device analysis revealed that the imaging window detached from the lapjoint area.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTICROSS¿
Type of Device
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS  
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5700615
MDR Text Key46614160
Report Number2134265-2016-04620
Device Sequence Number1
Product Code OBJ
Combination Product (y/n)N
PMA/PMN Number
K123621
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 05/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/28/2017
Device Model NumberH749518080
Device Catalogue Number51808
Device Lot Number18854427
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/28/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/09/2016
Initial Date FDA Received06/06/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/29/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-