(b)(4).A visual, functional and dimensional inspection of the product involved in the complaint could not be conducted since the product was not returned.The device history record investigation did not show issues related to complaint.A document assessment (fmea) was conducted and no changes required.Corrective actions cannot be established since it is necessary to receive the physical sample to perform a proper investigation and confirm the alleged defect.At this time since the device sample is not available, it is not possible to determine the source of the defect reported.Corrective actions cannot be established since it is necessary to receive the device sample to perform a proper investigation and confirm the alleged defect.Teleflex will continue to monitor and trend on similar complaints.
|
(b)(4).The sample was returned for evaluation.A visual exam was performed and there were no defects observed.Functional testing was performed and the unit was connected to 110vac.The unit passed the initial power connect test and navigated through the power-on self-test (p.O.S.T.) with no issues.During the temperature display accuracy test the unit displayed the correct temperatures and properly alarmed in the high temperature scenario.The unit was prepared for the functional bench test where water, an adult breathing circuit ((b)(4)), 10 lpm of air pressure, and a low compliance column is connected to the unit for a real time operational scenario.The unit successfully negotiated all pre-operational self-tests again and was functioning real time.The unit temperature was set at 37 degrees c.It achieved the set temperature and maintained that temperature for 45 minutes.The unit functioned without interruption.Based on the investigation performed, the reported complaint could not be confirmed.Functional testing did not reveal any operational anomalies.
|