• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA PADGETT MODEL S SLIMLINE ELECTRIC DERMATOME; SKIN GRAFT PRODUCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA PADGETT MODEL S SLIMLINE ELECTRIC DERMATOME; SKIN GRAFT PRODUCTS Back to Search Results
Catalog Number 3539700
Device Problems Break (1069); Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913); Noise, Audible (3273)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
It was reported the device stops working intermittently and there are also ¿sounds¿ coming from the hand piece.Event circumstance reported as ¿unknown.¿ ¿no injury was reported.¿.
 
Manufacturer Narrative
Integra has completed their internal investigation on 28jun2016.The investigation activities included: methods: evaluation of actual device.Review of device history records.Review of complaint history.Results: device history record reviewed for s-1467 manufactured on 02/27/2015 show no abnormalities related to reported incident found.This device passed all required inspection points with no associated mrr's, variances or rework.No manufacturing or design related trend has been identified.Conclusion: the damaged micro switch led that device stops working intermittently, root cause could be damaged component.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PADGETT MODEL S SLIMLINE ELECTRIC DERMATOME
Type of Device
SKIN GRAFT PRODUCTS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH 45227
Manufacturer Contact
maria leonard
311 enterprise drive
plainsboro, NJ 08536
6099362393
MDR Report Key5701259
MDR Text Key47896270
Report Number3004608878-2016-00131
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/10/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number3539700
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/10/2016
Initial Date FDA Received06/06/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/12/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/27/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-