ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Device Problem Loss of Power (1475)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The pump has been returned to animas.Evaluation has not yet been completed.When evaluation is complete a supplemental report will be filed.No conclusion can be made at this time.
 
Event Description
On (b)(6) 2016, the reporter contacted animas, alleging a power (no power) issue.Reportedly, there were no damages to the battery cap or the battery compartment.Allegedly, there was no moisture in the pump.This complaint is being reported because the user may be unaware that the pump has lost power, leading to under delivery.There was no indication that the product caused or contributed to an adverse event.
 
Manufacturer Narrative
Device evaluation: the device has been returned and evaluated by product analysis on 06/23/2016 with the following findings: investigators were unable to adequately investigate and confirm the original complaint due to an additional failure; upon pump power up, a call service alarm 069 was reproduced and was unable to be cleared post-reboot attempt.The call service alarm 069 failure was defined as a language file corruption upon the language text retrieval and was recorded in the black box data as a call service alarm 87 failure.The investigators determined that the error is resident to u39 flash chip failure.The battery cap was returned with the pump and was able to secure to the pump and to a test pump.The pump was opened and no damages were observed to the power circuit.There was no evidence of internal moisture.Unrelated to the original complaint, the battery compartment was cracked.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Type of Device
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer (Section G)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer Contact
karin sargrad
200 lawrence dr
west chester, PA 19380-3428
4843561808
MDR Report Key5703027
MDR Text Key46874555
Report Number2531779-2016-11168
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 05/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/07/2016
Was the Report Sent to FDA? Yes
Device Age17 MO
Initial Date Manufacturer Received 05/18/2016
Initial Date FDA Received06/07/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/21/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/02/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown