• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS FEM COMPONENT, CEMENTED; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS FEM COMPONENT, CEMENTED; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5515-F-501
Device Problem Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/12/2016
Event Type  malfunction  
Manufacturer Narrative
Only pacaging was returned, device was implanted.When completed, the investigation results will be submitted in a supplemental report.
 
Event Description
The #5 lt femoral component stuck in both sterile packages.Delay of surgery was 5 min.
 
Manufacturer Narrative
An event regarding packaging issue involving a triathlon ps fem component, cemented was reported.The event was not confirmed.Method & results: -device evaluation and results: the packaging box with a shrink wrap partially open, package insert, tyvek lid of the outer blister and the ifu were returned.The implant, tyvek lid of the inner blister and the foam was not returned.However, there is evidence that there was a seal in the inner and outer blisters.The packaging box and outer sterile packaging looks remarkable.The inner sterile packaging looks as if it was pushed inside out.-medical records received and evaluation: not performed as medical records were not provided for review.-device history review: a device history review confirmed all devices accepted into finished goods conformed to specification.-complaint history review: no other events were reported for the lot indicated.Conclusions: the exact cause of the event could not be determined because the packaging was returned without the implant.No further investigation for this event is possible at this time.If additional information becomes available, this investigation will be reopened.
 
Event Description
The #5 lt femoral component stuck in both sterile packages.Delay of surgery was 5 min.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PS FEM COMPONENT, CEMENTED
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
cindy chuhinko
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5710091
MDR Text Key48047109
Report Number0002249697-2016-01861
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2021
Device Catalogue Number5515-F-501
Device Lot NumberUIDYD
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/12/2016
Initial Date FDA Received06/09/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/12/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/18/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age77 YR
Patient Weight76
-
-