ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Device Problem Loss of Power (1475)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event Description
On (b)(6) 2016, the reporter contacted animas, alleging that the pump lost power.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the user may be unaware that the pump has lost power, leading to under delivery.
 
Manufacturer Narrative
Follow-up #1 date of submission 08/15/2016-device evaluation: the pump was returned and evaluated by product analysis on 07/19/2016 with the following findings: animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.A review of the pump black box data revealed no evidence of a power issue.During testing, the battery cap secured properly to the pump to maintain power.The rewind, load, and prime steps were completed with no duplicated power issues.The pump case was removed, and no intermittent conditions were found.The complaint was not duplicated.Unrelated to the complaint, the battery compartment was observed to be cracked.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Type of Device
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer (Section G)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
Manufacturer Contact
karin sargrad
200 lawrence dr
west chester, PA 19380-3428
4843561808
MDR Report Key5710519
MDR Text Key47296893
Report Number2531779-2016-11482
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 05/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2016
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Device Age31 MO
Initial Date Manufacturer Received 05/20/2016
Initial Date FDA Received06/09/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/15/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/18/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-